Model CPH62I0

Intrinsically-safe hand-held pressure indicator (ATEX version)

1-channel version or 2-channel version

Selected documents
Sales contact Questions? Find a contact in your vicinity.

Applications

  • Calibration service companies/service industry
  • Instrument and control workshops
  • Quality assurance

Special Features

  • Digital indicator with interchangeable pressure sensors (plug-and-play)
  • Measuring ranges from 0 ... 100 mbar to 0 ... 1,000 bar (0 ... 1.5 psi to 0 ... 14,500 psi)
  • Accuracy: 0.2 %, optional 0.1 % (incl. calibration certificate)
  • Intrinsically safe version, II 2G Ex ib IIC T4
  • Software and complete service cases (incl. pumps) available

Description

Extensive application possibilities
Stainless steel pressure sensors with measuring ranges up to 1,000 bar (14,500 psi) are available for the model CPH62I0 hand-held pressure indicator. It is therefore particularly suitable as a test instrument for applications such as process engineering, chemical industry, refineries, etc.

The digital indicator automatically detects the measuring range of the connected pressure sensor and guarantees a highly accurate pressure measurement.

Functionality
The CPH62I0 can be used for measuring both gauge and absolute pressure. Differential pressure measurement is possible with the 2-channel version CPH62I0-S2, and two connected model CPT62I0 reference pressure sensors. Selectable pressure units here are bar, mbar, psi, Pa, kPa, MPa, mmHg or inHg.

An integrated data logger and various other functions such as Min, Max, Hold, Tare, zero point adjustment, alarm, poweroff, peak value detection (1,000 measurements/s), average value filter, etc. ensure that the CPH62I0 can be used for many different applications.

 

Software
In addition to the GSoft data-logger evaluation software for the tabular and graphical representation of the logged data, WIKA-CAL calibration software for calibration tasks is also available. WIKA-CAL also offers, over and above PC-supported calibration, the management of the calibration and instrument data in an SQL database. For data transfer, an RS-232 and a USB interface are available.

Complete test and service cases
For maintenance and service applications, various service case systems are available. These include service cases with or without pressure generation, battery, connection adapter, etc.

Certified accuracy
For each reference pressure sensor, the accuracy for the complete measuring chain is certified by a factory calibration certificate which accompanies the instrument. On request, we can provide a DKD/DAkkS calibration certificate for this instrument.

Wish list - Documents (0)

There are currently no documents on your wish list. You can add any document from the download area to your wish list. To do this, select the desired language using the flag symbol.

The maximum number of 20 entries was reached.

To receive your download link via e-mail, enter your address here:

The e-mail address is invalid.